Additel's ADT221A is a highly integrated Multifunction Temperature Calibrator featuring several patented technologies, the ADT221A is an ultra-compact, rugged, and easy to use hand-held device for sourcing, simulating and measuring temperature, and electrical signals. With its smartphone-like menu and interface make the operation simpler and easier. The ADT221A is ideal for calibrating, maintaining, and troubleshooting process instruments. Automation and documentation capabilities make the ADT221A a turnkey solution.
Applications
The ADT221A multifunction temperature calibrator is a process tool for measuring, sourcing and simulating mA, mV, V, RTDs, thermocouples, ohms, frequency, and pulses, captures switch values and provides 24 V loop power.
Electrical
- Resistance measurement / simulation
- Voltage measurement / generation
- Current measurement /generation
- Frequency measurement / generation
- Pulse counting/generation
- 24 VDC loop supply
- Switch sensing
Temperature
- RTD measurement /simulation
- TC measurement /simulation
Features #twoj_fragment1-1
Electrical Specifications #twoj_fragment1-2
Specifications #twoj_fragment1-3
Temperature Specifications #twoj_fragment1-4
Order Info #twoj_fragment1-5
Video Demo #twoj_fragment1-6
Sourcing, simulating and measuring temperature and electrical signals
- Sources and measures mV, mA, ohms, RTDs, thermocouples, frequency, and pulses
- Simulates and measures 13 thermocouples and 11 RTDs to calibrate transmitters
- Measures and sources pressure using 160 series Intelligent Digital Pressure Modules from -15 psi to 10,000 psi (-1 bar to 700 bar)
- 24V loop power supply
- Simultaneous dual reading capability
- Automatic switch test
- Supports square root transmitter
- Pulse frequency output for the calibration of flow totalizer
Easy to use
- Smartphone-like menu and interface make the operation simpler and easier
- Ultra-compact, size 3.9” x 7.6” x 2.0” (100mm x 192mm x 52mm), and weight 1.6 lb (0.7 kg)
- One hand operation
Calibrated cold junction compensation (Patented)
- Cold junction equivalent block in the calibrator
- A calibrated PRT element with flexible leads is installed in the equivalent block for thermocouple cold junction compensation
- This PRT element can be pulled out from the calibrator and re-calibrated and corrected at ice point by users
Built-in temperature readout
- CVD coefficients of a calibrated PRT can be input into the calibrator for accurate temperature measurement
Multi lingual interface
- English, German, French, Italian, Spanish, Portuguese, Simplified Chinese (Traditional Chinese, Japanese and Russian are available per request)
Documenting and automated procedure capability
- Manage the information of the device under test.
- Set up automated calibration procedures, and ADT 221A performs the test, calculates the errors, displays and/or stores the results in the memory, and highlights the out-of-tolerance points.
- As-found and As-left functions allow recording and documenting results for quality control.
- Download tasks and upload the results.
- Snapshots allow you to capture and save work.
Build-in unit conversion tool
- Build-in converters for pressure units, temperature units, temperature vs. resistance (RTDs), and temperature vs millivolt (thermocouples)
Display 3.5 inch TFT color screen
Rugged
- Rugged design for harsh environments.
- Passed a 1-meter drop test.
- Three year warranty for the 221A, and one year for the battery pack
Misuse protection
- Up to 30V voltage on any two sockets and up to 1A current on current sockets will not damage the calibrator.
- The calibrator will return to normal condition as soon as the voltage or current is removed.
Rechargeable battery
- Rechargeable Li-ion battery for 15 hours uninterrupted use.
- Battery life will be reduced when 24V is applied.
- The rechargeable battery is replaceable.
NSIT Traceable Calibration with Data
Warranty 3 years
Electrical Measure Specifications
Measurement Accuracy
|
|
Range
|
Resolution
|
Accuracy
|
Voltage DC
|
±75.0000 mV
±30.0000 V
|
0.1µV
0.1mV
|
0.01%RD + 3.75 μV
0.01%RD + 1.5 mV
|
Current DC
|
±30.0000 mA
|
0.1µA
|
0.01%RD + 1.5 μA
|
Resistance
|
Two-wire
|
0 to 400.000 Ω
|
1mΩ
|
0.02%RD + 0.02 Ω
|
Three-wire
|
0 to 400.000 Ω
|
1mΩ
|
0.02%RD + 0.02 Ω
|
Four-wire
|
0 to 400.000 Ω
|
1mΩ
|
0.01%RD + 0.02 Ω
|
Two-wire
|
0 to 4000.000 Ω
|
10mΩ
|
0.02%RD + 0.2 Ω
|
Three-wire
|
0 to 4000.000 Ω
|
10mΩ
|
0.02%RD + 0.2 Ω
|
Four-wire
|
0 to 4000.000 Ω
|
10mΩ
|
0.01%RD + 0.2 Ω
|
Frequency
|
1 to 50000.0 Hz
|
0.1Hz
|
0.005%RD + 1 Hz
|
Pulse
|
0 to 999999
|
1
|
N/A
|
Limit Switch
|
If the switch has detective voltage, its range is +3V to +24V
|
Source Accuracy
|
Voltage DC
|
-10.000 to 75.000mV
0 to 12.0000 V
|
1μV
0.1mV
|
0.02%RD + 4.25 μV 0.02%RD + 0.6 mV
|
Current DC
|
0 to 22.000 mA
|
1µA
|
0.02%RD + 1.1 μA
|
Resistance
|
1 to 400.00 Ω
1 to 4000.0 Ω
|
10mΩ
100mΩ
|
0.02%RD + 0.02 Ω 0.03%RD + 0.4 Ω
|
Frequency
|
0 to 50000.0 Hz
|
0.1Hz
|
0.005%RD + 1 Hz
|
Pulse
|
0 to 999999
|
1
|
N/A
|
DC24V
|
N/A
|
N/A
|
0.5V
|
Evnironmental Specifications |
Operating temperature |
-10ºC to 50ºC |
Storage temperature |
-20ºC to 60ºC |
Humidity |
<90%, non-condensing |
Safety Specifications |
European compliance |
CE Mark |
Mechanical Specifications |
Display |
3.5 inch TFT colour screen |
Electrical connection |
φ4mm sockets and flat mini-jack thermocouple socket |
RS232 interface |
standard RS232-DB9 socket |
Size |
100 x 192 x 52mm |
Weight |
0.7kg |
Power supply |
Polymer Li-ion rechargeable battery, or 10V DC adaptor |
Battery |
Rechargeable Li-ion battery (included) |
Battery life |
15 hours uniterrupted use
Battery life will be reduced when 24V is applied
|
Battery charge |
110V/220V external power adaptor (included) |
Temperature Specifications
Thermocouple Measurement and Source accuracy
|
Measure
and Simulate
|
Standard
|
Temperature Range (°C)
|
Accuracy (°C)
|
Measure
|
Source
|
S
|
IEC 584
|
-50 to 1768
|
-50 to 400
|
1.0
|
1.1
|
400 to 1000
|
0.6
|
0.6
|
1000 to1768
|
0.7
|
0.8
|
R
|
IEC 584
|
-50 to 1768
|
-50 to 200
|
1.4
|
1.4
|
200 to 500
|
0.6
|
0.6
|
500 to 1768
|
0.6
|
0.7
|
B
|
IEC 584
|
0 to 1820
|
50 to 450
|
3.8
|
3.8
|
450 to 800
|
0.9
|
0.9
|
800 to 1820
|
0.6
|
0.7
|
K
|
IEC 584
|
-270 to 1372
|
-250 to -200
|
1.0
|
1.1
|
-200 to -100
|
0.4
|
0.5
|
-100 to 600
|
0.3
|
0.3
|
600 to 1372
|
0.4
|
0.5
|
N
|
IEC 584
|
-270 to 1300
|
-250 to -200
|
1.5
|
1.6
|
-200 to -100
|
0.5
|
0.6
|
-100 to 1300
|
0.4
|
0.5
|
E
|
IEC 584
|
-270 to 1000
|
-250 to -200
|
0.6
|
0.7
|
-200 to -100
|
0.3
|
0.3
|
-100 to 0
|
0.2
|
0.2
|
0 to 700
|
0.2
|
0.3
|
700 to 1000
|
0.2
|
0.4
|
J
|
IEC 584
|
-270 to 1200
|
-210 to -100
|
0.3
|
0.3
|
-100 to 1200
|
0.3
|
0.4
|
T
|
IEC 584
|
-270 to 400
|
-250 to -200
|
0.8
|
0.9
|
-200 to 0
|
0.4
|
0.4
|
0 to 400
|
0.2
|
0.2
|
C
|
ASTM E988
|
0 to 2315
|
0 to 1000
|
0.5
|
0.5
|
1000 to 1800
|
0.7
|
0.9
|
1800 to 2315
|
1.0
|
1.4
|
D
|
ASTM E988
|
0 to 2320
|
0 to 100
|
0.5
|
0.5
|
100 to 1100
|
0.4
|
0.5
|
1100 to 2000
|
0.6
|
0.9
|
2000 to 2320
|
0.9
|
1.3
|
G
|
ASTM E1751
|
0 to 2315
|
0 to 200
|
2.4
|
2.4
|
200 to 400
|
0.5
|
0.5
|
400 to 1400
|
0.4
|
0.5
|
1400 to 2315
|
0.7
|
1.0
|
L
|
DIN43710
|
-200 to 900
|
-200 to -100
|
0.2
|
0.3
|
-100 to 400
|
0.2
|
0.2
|
400 to 900
|
0.2
|
0.3
|
U
|
DIN43710
|
-200 to 600
|
-200 to 0
|
0.4
|
0.4
|
0 to 600
|
0.2
|
0.3
|
RTD Measurement and Source accuracy
|
Measure
and Simulate
|
Standard
|
Temperature Range (°C)
|
Accuracy (°C)
|
Measure (2W/3W)
|
Measure (4W)
|
Source
|
Pt10(385)
|
IEC 751
|
-200 to 850
|
-100 to 200
|
0.65
|
0.60
|
0.65
|
200 to 600
|
0.82
|
0.72
|
0.82
|
600 to 850
|
0.96
|
0.82
|
0.96
|
PT100
(385)
|
IEC 751
|
-200 to 850
|
-100 to 200
|
0.15
|
0.1
|
0.15
|
200 to 600
|
0.26
|
0.16
|
0.26
|
600 to 850
|
0.34
|
0.20
|
0.34
|
Pt100
(3916)
|
IEC 751
|
-200 to 850
|
-100 to 200
|
0.15
|
0.1
|
0.15
|
200 to 600
|
0.26
|
0.16
|
0.26
|
600 to 850
|
0.33
|
0.20
|
0.33
|
Pt200(385)
|
IEC 751
|
-200 to 850
|
-100 to 200
|
0.37
|
0.32
|
0.69
|
200 to 600
|
0.51
|
0.41
|
0.92
|
600 to 850
|
0.61
|
0.48
|
1.08
|
Pt500(385)
|
IEC 751
|
-200 to 850
|
-100 to 200
|
0.20
|
0.16
|
0.36
|
200 to 600
|
0.32
|
0.22
|
0.54
|
600 to 850
|
0.40
|
0.27
|
0.67
|
Pt1000
(385)
|
IEC 751
|
-200 to 850
|
-100 to 200
|
0.1
|
0.05
|
0.26
|
200 to 600
|
0.2
|
0.10
|
0.42
|
600 to 850
|
0.27
|
0.14
|
0.54
|
Cu100(428)
|
GOST6651-94
|
-50 to 150
|
-50 to150
|
0.12
|
0.09
|
0.12
|
Cu50(428)
|
GOST6651-94
|
-50 to 150
|
-50 to 150
|
0.17
|
0.13
|
0.17
|
Cu10(427)
|
Minco Application Aid #18
|
-100 to 260
|
-100 to 260
|
0.61
|
0.56
|
0.61
|
Ni120(672)
|
Edison curve #7
|
-100 to 260
|
-100 to 260
|
0.07
|
0.05
|
0.07
|
Ni100(618)
|
DIN 43760
|
-100 to 260
|
-100 to 260
|
0.08
|
0.06
|
0.08
|
Model Number: ADT221A
Accessories (included)
110V/220V external power adapter
|
1 pc
|
Chargeable Li-ion battery
|
1 pc
|
Test leads
|
2 sets (4 pcs)
|
Short circuit cable
|
1 sets (2 pcs)
|
Manual
|
1 pc
|
NIST traceable calibration certificate
|
1 pc
|
Optional Accessories
Model number
|
Description
|
9050
|
USB to RS232 (DB-9 Male) Adapter
|
9080
|
Cold Junction compensation kits (including TC plug, compensation cable, S,R,B,K,J,T,E,N)
|
9712
|
Spare chargeable Li-ion battery for multifunction calibrator
|
9816
|
110V/220V external power adapter for multifunction calibrator
|
9906
|
Carrying case for multifunction calibrator
|
9510
|
Additel/Cal Task management software for multifunction calibrator
|